题名 | Highly sensitive and miniaturized microcone-curved resonant photoacoustic cavity for trace gas detection |
作者 | |
通讯作者 | Ni, Wenjun |
发表日期 | 2024-12-01
|
DOI | |
发表期刊 | |
ISSN | 2213-5979
|
卷号 | 40 |
摘要 | This paper proposes a novel microcone-curved resonant photoacoustic cell (MCR-PAC) for highly sensitive trace gas detection. The MCR-PAC features with microcone-curved resonant region and cylindrical buffer chamber, which dominates the photoacoustic signal amplification. By introducing the hyperbolic eccentricity as a new optimization dimension, the quality factor of the MCR-PAC is remarkably strengthened to enhance the acoustic pressure amplitude. At an eccentricity value of 5, the volume of the photoacoustic resonant cavity is approximately 0.23 cm(3). Targeting trace acetylene, the system achieves a minimum detection limit of 1.41 ppb with an integration time of 290 s, corresponding normalized noise equivalent absorption coefficient is 1.88x10(-9) W |
关键词 | |
相关链接 | [来源记录] |
收录类别 | |
语种 | 英语
|
学校署名 | 其他
|
资助项目 | National Natural Science Foundation of China["62105373","62171487"]
; Natural Science Foundation of Hubei Province["2023AFC012","2024AFA030"]
; Knowledge Innovation Program of Wuhan-Shuguang Project[2022010801020408]
|
WOS研究方向 | Acoustics
; Engineering
; Instruments & Instrumentation
; Radiology, Nuclear Medicine & Medical Imaging
|
WOS类目 | Acoustics
; Engineering, Biomedical
; Instruments & Instrumentation
; Radiology, Nuclear Medicine & Medical Imaging
|
WOS记录号 | WOS:001319788500001
|
出版者 | |
来源库 | Web of Science
|
引用统计 | |
成果类型 | 期刊论文 |
条目标识符 | http://sustech.caswiz.com/handle/2SGJ60CL/834340 |
专题 | 工学院 工学院_电子与电气工程系 |
作者单位 | 1.South Cent Minzu Univ, Coll Elect & Informat Engn, Hubei Engn Res Ctr Intelligent IOT Technol, Hubei Key Lab Intelligent Wireless Commun, Wuhan 430074, Peoples R China 2.Huazhong Univ Sci & Technol, Sch Opt & Elect Informat, Wuhan Natl Lab Optoelect WNLO, Wuhan 430074, Peoples R China 3.Huazhong Univ Sci & Technol, Natl Engn Res Ctr Next Generat Internet Access Sys, Sch Opt & Elect Informat, Wuhan 430074, Peoples R China 4.Southern Univ Sci & Technol, Dept Elect & Elect Engn, Coll Engn, Shenzhen 518055, Peoples R China |
推荐引用方式 GB/T 7714 |
Zhao, Zhongke,Ni, Wenjun,Yang, Chunyong,et al. Highly sensitive and miniaturized microcone-curved resonant photoacoustic cavity for trace gas detection[J]. PHOTOACOUSTICS,2024,40.
|
APA |
Zhao, Zhongke.,Ni, Wenjun.,Yang, Chunyong.,Ran, Sixiang.,He, Bingze.,...&Shum, Perry Ping.(2024).Highly sensitive and miniaturized microcone-curved resonant photoacoustic cavity for trace gas detection.PHOTOACOUSTICS,40.
|
MLA |
Zhao, Zhongke,et al."Highly sensitive and miniaturized microcone-curved resonant photoacoustic cavity for trace gas detection".PHOTOACOUSTICS 40(2024).
|
条目包含的文件 | 条目无相关文件。 |
|
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论